Stanford SR630 Thermocouple Monitor

16 channel thermocouple monitor

Stanford Research

Distributed only in Germany
This is a discontinued product
More Information? Please contact us.
  • Overview

Key Facts

  • 16 channels
  • B, E, J, K, R, S and T type thermocouples
  • 0.1 °C resolution
  • Displays °C, °F, K and VDC
Down to Key Facts

Brief Description

The SR630 is a 16-channel thermocouple monitor designed to read, scan, print and log temperatures or voltages. You can use any one of seven standard thermocouple types to read temperatures from -200 °C to +1700 °C.

Features & Benefits

Sixteen screw-terminal inputs are mounted on a rear-panel isothermal block for cold junction compensation.

Specifications

Thermocouple
Channels 16
Thermocouple types B, E, J, K, R, S, T
Down to Specifications

Key Facts

  • 16 channels
  • B, E, J, K, R, S and T type thermocouples
  • 0.1 °C resolution
  • Displays °C, °F, K and VDC
  • 2,000 point non-volatile memory
  • Four analog outputs proportional to temperature
  • GPIB, RS-232 and printer interfaces

Brief Description

Temperature Monitor

The SR630 is a 16-channel thermocouple monitor designed to read, scan, print and log temperatures or voltages. You can use any one of seven standard thermocouple types to read temperatures from -200 °C to +1700 °C. For remote monitoring applications, the SR630 can time-stamp and store up to 2000 readings in non-volatile memory for later analysis.

Temperature readings from the SR630 can be viewed on the front panel or queried via the instrument’s standard RS-232 or GPIB interfaces. In addition, the standard Centronics printer port provides convenient hardcopy output in either tabular or strip-chart format.

Voltage Monitor

The SR630 can also be configured as a 16-channel DC voltmeter with full-scale ranges from 30 mV to 100 V (1 mV resolution). As a voltmeter, the unit has 0.05 % accuracy and 1 µV input offset drift. Each channel can be set independently to monitor either temperature or voltage, giving the SR630 even more flexibility in your application. Four rear-panel outputs are also available to provide a voltage proportional to the temperature of the first four input channels. The voltage outputs can be used to drive recorders or to control external instrumentation.

Features & Benefits

  • Inputs

    Sixteen screw-terminal inputs are mounted on a rear-panel isothermal block for cold junction compensation. The isolated differential inputs have a 250 V breakdown level, allowing the SR630 to tackle difficult applications such as temperature profiling of electrically live equipment. Each of the 16 channels may be independently set to display in units of °C, °F, K, mV or V. Similarly, thermocouple type, nominal temperature, temperature limit, and alarms may be uniquely set for each channel, providing complete flexibility in the configuration of the instrument. Access to any channel parameter is provided through the front panel or via the computer interfaces. The SR630 can store up to nine entire instrument configurations including thermocouple type and temperature limits for all 16 channels.

  • Outputs

    There are four analog outputs that are proportional to the readings on channels one through four. Both the slope (m) and offset (b) of the equation, Vout = ±mX + b, are changed by setting the nominal temperature and chart span for the corresponding channel. X is either temperature or voltage, depending on the setting for each channel.

    You may use the analog outputs to drive chart recorders or they can provide a feedback signal for proportional temperature control systems. They may also be used as fixed-level, analog control signals (set via the computer interfaces).

  • Alarms

    Each channel can be programmed individually with an upper and lower temperature limit. An audio alarm and a relay closure indicate when any channel exceeds its preset temperature or voltage limits. The front-panel LED display will indicate which channel has exceeded its limit.

  • Data Logging

    An internal, battery backed-up clock/calendar is used to time-stamp temperature readings. The SR630 can be set to scan, time-stamp, and record data on up to 16 channels at intervals from 10 to 9999 seconds. Data is either sent immediately to the printer, or stored in the internal 2,000 point buffer for output at a later time (via GPIB or RS-232).

  • Built-In Printer Interface

    A standard Centronics printer interface makes it simple to get hardcopy output of temperature or voltage scans. Two hardcopy formats are provided. The SR630 can print a continuous strip chart showing up to 16 different temperatures. In addition, data can be printed in a tabular format which logs the time, date, and temperature or voltage for each channel.

  • Analog Multiplexer

    The SR630 can also function as a 1:15 analog multiplexer. Any of the first 15 input channels can be switched out to channel 16 and passed on to other instruments. This feature is useful in ATE systems and many other monitoring applications.

  • Available Thermocouples

    Although the SR630 will read types B, E, J, K, R, S and T type thermocouples, for many applications type K, or Chromel/Alumel, will serve well. K-type thermocouples offer a wide temperature range (-200 °C to +1250 °C), a low standard error, and good corrosion resistance.

Specifications

Thermocouple
Channels 16
Thermocouple types B, E, J, K, R, S, T
Display units °C, °F and K
Display resolution 0.1 °C
Temperature displays Actual, Nominal, or Offset
Accuracy 0.5 °C for J, K, E and T. 1.0 °C for R, S and B. Errors are for the SR630 only. Standard errors for thermocouple wire are 2 to 5 times the error due to the SR630.
Inputs
Channels 16
Input type Independent, floating and differential
Input resistance 10 MΩ between + and - terminals, >1 GΩ to ground
Input capacitance 0.001 µF
Input bias current <100 pA
Input protection 250 Vrms
Open circuit check 250 µA
Common mode ±200 VDC
Voltmeter
Full-scale display (±9.999, ±99.99 or ±999.9) mVDC, (±9.999 or ±99.99) VDC
Range select Automatic
Resolution ±1 of least significant displayed digit
Offset ±2 of least significant displayed digit
Gain accuracy 0.05 %
Conversion rate 10/s for 50 Hz line, 12/s for 60 Hz
Line rejection >100:1
Scanning and Data Logging  
Scanning Selected channels will be scanned. Dwell time between scans is set from 10 s to 9999 s.
Alarm Temperature or voltage limit for each channel
Scan enable All channels may be scanned or skipped
Printer output Voltages, Temperatures, Time and Date as a list or in a graphical format
Data memory Last 2000 measurements in battery backed-up memory
General
Analog outputs Four analog voltage outputs proportional to temperature of channels 1, 2, 3 and 4
Relay output Switching 10 A
Store and recall Nine locations for instrument set-up
Interfaces RS-232, GPIB, and Centronics ports (standard). All instrument functions may be set and read via RS-232 or GPIB.
Power 10 W, 100/120/220/240 VAC, 50/60 Hz
Dimensions 8.5" × 3.5" × 13" (WHL)
Weight 9 lbs.
Warranty One year parts and labor on defects in materials and workmanship
Thank you.

Your request was successfully sent!

We will get in touch with you soon.

Your Rohde & Schwarz Service Team

Feedback
Tell us your opinion
Please tell us which topic.
Please enter your inquiry.
Contact information

Please enter your first name.

Please enter a valid e-mail address.

Please enter your name.
Security Response Code
show captcha

The input was not correct.
*Mandatory Field
Contact Us
Request Information
Thank you.

Your request was successfully sent!

We will get in touch with you soon.

Your Rohde & Schwarz Service Team

Please enter your inquiry.

Please enter your first name.

Please enter your name.

Please enter a valid e-mail address.

Please enter your company name.

Please enter your city name.

Please enter your postal code.

Please enter your country.
show captcha
The input was not correct.
*Mandatory Field
Request a Quote
Thank you.

Your request was successfully sent!

We will get in touch with you soon.

Your Rohde & Schwarz Service Team

Please enter your inquiry.

Please enter your first name.

Please enter your name.

Please enter a valid e-mail address.

Please enter your phone number.

Please enter your company name.

Please enter your city name.

Please enter your postal code.

Please enter your country.
show captcha
The input was not correct.
*Mandatory Field
Request a Demo
Thank you.

Your request was successfully sent!

We will get in touch with you soon.

Your Rohde & Schwarz Service Team

Please enter your inquiry.

Please enter your first name.

Please enter your name.

Please enter a valid e-mail address.

Please enter your company name.

Please enter your city name.

Please enter your postal code.

Please enter your country.
show captcha
The input was not correct.
*Mandatory Field

Downloads
International Website